WIRING FORMATS (cont’d)
BIC-II // CompuScale IIA / CompuFlo Interconnection
1. BIC-II, ports 1-6, Rx input is non-polarized.
2. Refer to instruction manual PL-276 and PL-281 for wiring of CompuFlo and CompuScale IIA respectively.
3. BIC-II interconnecting cable shield should be grounded at CompuFlo and CompuScale IIA only. Insulate
shields at junctions to prevent inadvertent grounding. Use Belden cable 9552, 2 pair shielded,
18 AWG or equivalent. Maximum communication loop length is 3000 m ( 10,000 ft ).
4. All wiring, whether in non-hazardous (as shown) or hazardous areas, must be done in conjunction with
approved conduit, boxes and fittings, and to procedures in accordance with all governing regulations.
FIG. 3E
to other products
as required
RATE
4 - 20 mA
CONTLR
4 - 20 mA
SIG. A
SIG. B
SIG. COM.
EXC.
SEN.
SEN.
EXC.
SPD. EXC.
SPD. SIG
SPD. COM
CONT. 8
CONT. 7
CONT. 6
CONT. 5
CONT. 4
CONT. 3
CONT. 2
CONT. 1
TRANS
20 mA
REC
20 mA
DISP. COM.
CONT. COMMON
1TB
1
RELAY
34
33
32
31
30
29
28
27
26
25
24
23
22
21
20
19
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
PORT6PORT 5PORT 4
PORT 3
PORT 2
+
Rx
- +
Tx
-
+
Rx
- +
Tx
-
+
Rx
- +
Tx
-
+
Rx
- +
Tx
-
+
Rx
- +
Tx
--
Tx
+-
Rx
+
TB1
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
PORT 1
Board B
-
+
-
+
+
+
+
+
-
-
-
-
-
+
-
BIC-II
CompuScale IIA / CompuFlo
PL-371 12 – 7
Vue de la page 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 76

Commentaires sur ces manuels

Pas de commentaire