TROUBLESHOOTING
SELF TEST
The BIC-II is capable of performing two levels of automatic self diagnostic tests upon initial power up; Normal,
and Test. The on board LED indicators will illuminate in a specific sequence during these tests.
Should the LED’s not illuminate once the BIC-II, board B, power switch SW1 is flipped ON, check the
following :
› power has been supplied to board B, TB3-2/3
› the power selection jumper J2 is in the appropriate position
› the power fuse is the proper rating and in intact
Should an internal fault occur during one of these tests, LED 1-8 will flash red simultaneously three times and
then one LED will remain red, indicating the following :
LED1 UART failure
LED2 EPROM failure
LED3 RAM failure
LED4 EEPROM failure
LED5 Configuration Error
Return the BIC-II to Milltronics or your distributor for repair.
A self test will also be performed if the BIC-II receives a system test request message from the host computer
during normal operation. The results of this test will be returned to the computer, refer to PROTOCOL / BIC-II
Commands.
NORMAL
The BIC-II performs a series of microprocessor and communication circuitry tests.
Set the BIC-II board A, switch SW1 contact 8 to the NRM position. Flip the BIC-II, board B, POWER switch to
ON. LED1-8 all illuminate orange simultaneously. As the 8 communication ports are activated, the respective
LED is turned OFF, starting with the "HOST" (LED1), then "MBUS" (LED2), and followed by each of the 6
product ports (LED3-8).
After a successful test the BIC-II will automatically commence normal operation.
PL-371 10 – 1